Contact Us

Looking for a Product? Have any questions? Reach out to Us. We will get back to you within a day!


California
United States

+1 (510) 899-1340

Service AI specializes in providing accessories and spare parts for laboratory instrumentation in a niche area of the market. Our business strategy focuses on meeting the unique needs and requirements of popular equipment that is widely used but no longer manufactured or supported by its original manufacturers.   With over 25 years of experience in supporting scientists with laboratory instrumentation and spare parts, our knowledgeable and dedicated team takes pride in ensuring that every recertified or new instrument and spare part we supply is fully tested and certified by a manufacturer-trained engineer. We guarantee that our products will work and provide proof of performance documentation when applicable.  

We only supply refurbished laboratory instrumentation that our team of experts are trained and certified to work on, ensuring that we deliver high-quality instruments and reliable service to our customers.

At Service AI, we understand the importance of safe shipping and delivery, and we take utmost care in ensuring that our products are delivered safely to our clients. Additionally, we offer professional installation services upon request.

Lab Pundit

Service AI offers a comprehensive range of spare parts and fully certified and rigorously tested laboratory instruments, accompanied by detailed proof of performance documentation.

Explore Ask LabPundit for answers to commonly asked questions about our laboratory equipment. Should you have further inquiries, please don't hesitate to contact us via email.

Filtering by Tag: Beckman LS6500 troubleshooting

About LS6500 Racks

Service AI

The Beckman LS6500 uses racks for holding sample vials for counting. There are mainly two types of racks available for counting samples.

The standard (White) rack hold 12 standard size vials per rack. A maximum of 28 racks can be loaded on to the LS6500 work deck built with a robust rack sample changer at a time, which works out to 336 standard vials at a time.

The miniature (Blue) rack hold 18 miniature vials per rack. The user can load a maximum of 36 racks at a time, which works out to (18 x36=648) miniature vials at a time.

Another type of rack, which is rarely used and seen in the field is the Bio-Vial (Green) Rack. These racks hold 18 Bio-Vials and can only be used on an instrument that uses miniature racks.

A majority of LS6500 comes with the versa rack option installed in it that allows use of standard (White) racks as well as the miniature (blue and green) racks. If the LS6500 does not have the versa-rack option available, it can take only standard racks.

In an LS6500 that has the Versa-rack option available, you will find a separate Mini/Maxi sensor PC board (PN. 500726) mounted on the right rear changer panel of the instrument. This board will be mounted to the right side of the rack reader board (PN. 500728)

 About Color Coded Racks:

The LS6500 also comes with a red rack and yellow rack for use as a Halt Rack and Interrupt rack respectively. The color is for easy recognition ONLY. The user has to use these rack with the correct Command card to work!

Is your Beckman Coulter LS6500 failing efficiencies?

Service AI

Check the following:

  1. Inspect the sample vials to ensure they are securely sealed.

  2. Ensure the instrument is calibrated and set up properly.

  3. Verify that you have accurately corrected for decay using the Half-life Tables, and that your computations are correct.

  4. Replace LS6500 Elevator Cap.

  5. Inspect the detector reflector housing. If it looks dirty, replace the reflector housing.

Read More

Resolving the 'H# Abort: Count Rate Too Low' Error on Your LS 6500

Service AI

Even though the H# value is counted by the Cs137 source drive, the cause for this failure could be because of a problem with the front end of the electronics, which consists of the PMTs, pre-amp housing assembly, or the high voltage power supply. Of course the problem could be with the Cs137 source drive itself as well.